Two-channel digital Oscilloscopes 93
Technical characteristics
Inputs
DOX2025B
DOX2070B
DOX2100B
Input coupling
AC, DC, GND
Input impedance
1 MΩ ±2% || 18 pF ±3 pF
Max. voltage at input
300 V (DC+AC Pk) 300 Vmax CAT II
Ch to Ch insulation
(both channels with the
same V/div configuration)
>100:1 to 12.5MHz
>100:1 to 35MHz
>100:1 to 50MHz
Probe attenuator
1X, 10X
Probe attenuator
configuration factors
0.1x,0.2x,0.5x,1x,2x,5x,10x,20x,50x,100x,200x,500x,1000x,2000x,5000x,10000x
Vertical system
DOX2025B
DOX2070B
DOX2100B
Vertical sensitivity
2mV/div. -10 V/div. (order 1-2-5)
Range of variation of the
vertical position
2mV - 200mV: ±1.6V
206mV - 10V: ±40V
Vertical resolution
8 bit
Number of Channels
2
Analog bandwidth
25MHz
70MHz
100MHz
Single bandwidth
25MHz
70MHz
100MHz
Flatness of the Bandwidth
on the BNC input
DC-10% of the nominal BW: ±1dB
10%-50% of the nominal BW: ±2dB
50%-100% of the nominal BW: +2dB/-3dB
Low frequency limit to -3dB
with AC coupling
10Hz
Noise: Pk-Pk for a 3K
recording
0.6 div. for a mean of 10 Pk-Pk readings, 2mV to 10V/div ranges
0.7 div. for a mean of 10 Pk-Pk readings, with fine adjustment of the range
SFDR with the harmonics
(measurement by FFT)
35dB
Accuracy of DC gain for
a signal of amplitude 6 div.
± 4.0%: 5mV/div. to 10 V/div. in fixed gain ranges
± 5.0%: 2mV/div. variable gain ranges
Accuracy of the DC
measurements:
all ranges: ≤100mV/div.
± [3% * (|measurement| +|offset|) +1% * |offset| + 0.2 div. +2mV]
Accuracy of the DC
measurements:
all ranges: > 100mV/div.
± [3% * (|measurement| +|offset|) +1% * |offset| + 0,2 div. +100mV]
Rise time
< 7.0ns
< 5.0ns
< 3.5ns
Overshoot (Typical)
(with pulse tr = 500ps)
<10% with probe or BNC input with 50 Ohm through termination
Skew between channels
(same V/div range)
< 4ns
< 4ns
< 1ns
Math operation
+, - , *, /, FFT
FFT
Type of Window: Hanning, Hamming, Blackman, Rectangular
Number of sampling points: 1024
Bandwidth
20MHz ±40%
(Note: The Bandwidth is limited to less than 10MHz when a probe is used at X1)
The oscilloscope must have operated continuously for thirty minutes at the service
temperature. You must run the Auto-calibration from the "Utility" menu if the service
temperature varies by more than 5°C. The oscilloscope must be within the factory
calibration range. All specifications are guaranteed except those marked "typical"