Functional description
Two-channel digital Oscilloscopes 83
Functional description
VIII - UTILITY system 3/4 (cont'd)
Option
Values
Instruction
Save
Store the mask configuration.
Load
Restore the mask configuration.
Return
Return to the main mask configuration menu.
Last page
Page 2/2
Return to the first page of the "Mask Configuration"
menu.
1. Press the UTILITY key to access the "Utility" menu.
2. Press the "Page after 1/4" key.
3. Press the "Page after 2/4" key to access the third page of the "Utility" menu.
4. Press the "Pass/Fail" key to access the "Pass/Fail" menu.
5. Press the "Activate test" key and choose "On"
6. Press the "Source" key to choose the source channel for the test.
7. Press the "Page after 1/2" key to access the second page of the "Utility" menu.
8. Press the "Mask Configuration" key to access the first page of the "Mask" menu.
9. Press the "X Mask" key and turn the "Adjust" knob to adjust the horizontal value
of the allowed range.
10. Press the "Y Mask" key and turn the "Adjust" knob to adjust the vertical value of
the allowed range.
11. Press the "Create mask" key to create the mask. You can also access the next
page of the "Mask" menu to restore the saved mask.
12. Go to the second page of the "Pass/Fail" menu. Press the "Stop On" key to
configure the option to Stop on Pass or Fail.
13. Go to the first page of the "Pass/Fail" menu. Press the "Operate" key to run
the pass/fail test.