Manual

Functional description
Two-channel digital Oscilloscopes 19
Functional description
I - VERTICAL system (cont'd)
1. Use the "Vertical position" knobs to move the traces up or down on the
screen.
2. While you adjust the vertical position of the trace, the value "Volts Pos = " is
displayed at bottom left on the screen.
3. Press the "Vertical position" knob to reset the vertical position to zero.
1. Use the "Volts/div" knobs to adjust the vertical sensitivity. If you turn the
"Volts/div" knob clockwise or anticlockwise, the oscilloscope increases or
reduces the sensitivity.
2. You can press the "Volt/div" to alternate between "Coarse" and "Fine"
adjustments. In the "Coarse" mode the vertical sensitivity varies in 1-2-5-
staged hops. In the "Fine" mode the vertical sensitivity varies continuously.
The reference command stores the traces. The reference function is available
once a source trace has been stored.
Option
Values
Instruction
Source
CH1 CH2
Choice of trace to be stored.
REFA
REFB
Choice of the location of storage or
restoration of a trace.
Save
Stores a source trace at the chosen location.
REFA/REFB
On
Off
Recalls the reference trace to the screen.
Deactivates the reference trace.
REF A β€œOn” REF A β€œOff ”
1. Press the "REF" button to access the REFerence menu.
2. Press the "Source" button to choose the source channel, CH1 or CH2.
3. Turn the vertical "POSITION" and "Volt/div" knobs to adjust the vertical
position and sensitivity to suitable values.
4. Press the "REF A" (or REF B) button to choose the storage location.
5. Press the "Save" button.
6. Select REF A (or REF B) "ON" to display the reference trace on screen.