Functional description
Two-channel digital Oscilloscopes 17
Functional description
I - VERTICAL system (cont'd)
The vertical scale can be adjusted in Coarse and Fine mode; the vertical
sensitivity band is from 2mV/div to 10V/div.
For example, for CH1:
Pressing "CH1""Volts/ Div" "Coarse" Volts/div, the default value.
The vertical sensitivity can be adjusted in hops in a 1-2-5 staging from
2mV/div, 5mV/div, and 10mV/div up to 10V/div.
Pressing "CH1""Volts/Div""Fine" Volts/div. This parameter allows
continuous adjustment of the vertical sensitivity.
From the Channel menu, select the probe coefficient corresponding to the
attenuation factor of the probe. To compensate a probe attenuation factor of
10:1, the probe coefficient must be set to 10X, in order to incorporate the
attenuation factor of the probe in the vertical sensitivity.
Coarse
Probe
attenuation
factor