Manual

Functional description
14 Two-channel digital Oscilloscopes
Functional description
AUTOSET (cont'd)
Function menu
Values
Acquisition
mode
Sample
Display format
Y-T
Type of display
Dots for a video signal; Vectors for an FFT spectrum;
otherwise unchanged
Vertical
coupling
DC or AC depending on the input signal
Bandwidth limit
Bandwidth limiter inactive (full band)
V/div
Adjusted
VOLTS/DIV
adjustment
Coarse (in hops, 1 2 5 staging)
Signal inverted
Inactive
Horizontal
position
Centre
S/div
Adjusted
Type of trigger
Edge
Source of
trigger
Autodetection of the input channel signal
Trigger edge
Rising
Trigger mode
Auto
Trigger
coupling
DC
Hold-Off
Minimum
Trigger level
Adjusted to 50%
When it leaves the factory, the oscilloscope is configured for normal use; this
is the "default configuration". To restore this configuration, press the
"DEFAULT SETUP" button. The parameterizing of the options, buttons, and
commands may be modified when you use the "DEFAULT SETUP" button;
see appendix B.
The DEFAULT SETUP button does not modify the following parameters:
Language option
Saved reference signal files
Saved configuration files
The display contrast
The calibration data