Operation Manual

MD 9225 Clamp Meter Series Specifications
17
ACA Current (Clamp-on)
RANGE Accuracy
1)
400.0A
40Hz ~ 60Hz @ 0 ~ 50A 1.0% + 6d
60Hz ~ 400Hz @ 0 ~ 50A 1.5% + 5d
40Hz ~ 60Hz @ 50A ~ 200A 1.5% + 5d
60Hz ~ 200Hz @ 50A ~ 200A 2.0% + 5d
40Hz ~ 60Hz @ 200A ~ 300A 2.0% + 5d
40Hz ~ 60Hz @ 300A ~ 400A 2.5% + 5d
1)
Induced error from adjacent current-carrying conductor: < 0.01A/A
True RMS Crest Factor:
< 1.8 : 1 at full scale & < 3.6 : 1 at half scale
Hz Frequency
Function Sensitivity (Sine wave) Range
400.0mV 350mV 10Hz ~ 1kHz
4.000V 3.2V 5Hz ~ 20kHz
40.00V 25V 5Hz ~ 100kHz
400.0V 100V 5Hz ~ 100kHz
600V 410V 5Hz ~ 5kHz
DCA/ACA Unspecified
Display counts: 5000
Maximum resolution: 0.001Hz
Accuracy: 0.5%+4d
Type-K Temperature
RANGE Accuracy
-20
o
C ~ 300
o
C 2.0% + 3d
300
o
C ~ 537
o
C 3.0% + 3d
-4
o
F ~ 572
o
F 2.0% + 6d
572
o
F ~ 1000
o
F 3.0% + 6d
Type-K thermocouple range & accuracy not included