Installation Guide
B1-MT400
15
Key 3- "OK" key: When you press the button, the device will perform calibration and the system will
automatically save parameters and return to the setting interface;
Key 4 - "EXIT" key: When you press the button, the system directly returns to the setting interface.
Note: This device supports two-point O
2
calibration, namely 0% and 21%.
3.7.6 "5.SET VOC"
Note: This option is used for VOC calibration;
Select "5.SET VOC" on the setting menu, and then press the "OK" button (Key 3) to enter the setting
interface. As shown in the following figure:
Operation Instructions of Buttons:
Key 1- "+" key: Press this button to add the calibration value by 0.001 ppm and press and hold this button
to add the calibration value by 0.01 ppm;
Key 2- "-" key: Press this button to reduce the calibration value by 0.001 ppm and press and hold this
button to reduce the calibration value by 0.01 ppm;
Key 3- "START" key: Press the button to start the calibration and display the calibration results: "OK"
indicates that the calibration is successful and "ERR" indicates that the calibration fails;
Key 4 - "EXIT" key: When you press the button, the system directly returns to the setting interface.
Note: The VOC calibration shall be performed by professional engineers. Users are not allowed to
perform such calibration by themselves.
3.7.7 "6.SET DATA FREQ"
Note: This option is used for setting data upload frequency;
Select "6.SET DATA FREQ" on the setting menu, and then press the "OK" button (Key 3) to enter
the setting interface. As shown in the following figure: