Installation Manual

33
probe connection wire is normal or broken,
otherwise replace the probe
Insert the probe into
another normal terminal device to check whether the
temperature can be normally uploaded, otherwise, it
means that the probe is faulty and needs to be
replaced.
Check whether the
MT500 connected with the PC is in a normal state.
After the MT200LM
terminal is locked or
unlocked, the corresponding
lock status information on
the monitoring software is
not consistent
Double click to the
lock status information of the terminal device to
view whether it can be synchronized.
Check whether the
antenna of the device is damaged; if the antenna is
damaged, it will affect the strength of the wireless
signal and result in data loss.
Check whether the
antenna on the terminal device is working properly,
and there is no metal around to block the signal.
Check for software
crash, and restart the software and try again.
7 Precautions
When adding liquid nitrogen, take care and prevent liquid nitrogen from
splash into MT200LM;
Do not use it in the environment below 0 for a long time.
The transmission distance of MT200LM is limited and should generally not exceed 3 rooms. If the
walls of the rooms are made of metal materials, it is better to install a data receiver in each room;
When the battery level of MT200LM is lower than 25%, the power consumption will be faster. Please
charge in time for the fear that the low-voltage protection of the battery acts and affect the use of the
equipment;
MT200LM is not waterproof, so the equipment shall not directly contact with liquid water or liquid