Installation Manual

21
3) Status Area Instructions of Liquid Nitrogen Tank
Detailed diagram of the
status area of liquid nitrogen
tank is shown in the main
interface.
Remarks:
a) Display the tank name which can be defined by user;
b) The SN of the intelligent lock installed on the tank;
c) Display the liquid nitrogen temperature detected by the temperature sensor (the temperature
measurement range of the temperature sensor is from -200 to +40 ; if no sensor is
connected, no sensor is displayed; if the temperature is higher than -190, a red message is
displayed. The default data upload frequency is 3 minutes per time). In addition, double-click
here, you can immediately query the current temperature of the device;
d) Display the battery level of the intelligent lock installed on the tank (the data upload frequency
and the temperature are consistent);
e) Display the switch status of the intelligent lock installed on the tank.
f) If N100LM normal temperature probe is equipped, the temperature difference will be displayed
here;