HCLH/SOP-YF-072-103 MT200LM Liquid Nitrogen Tank Monitoring Series Operation and Installation Manual
A1-MT200LM [Manufacturer]: Wuhan Huchuang Union Technology Co., Ltd. [Production Address]: No.1 Workshop, 1F, Building B10, Wuhan Hi-Tech Medical Device Park, No. 818 Gaoxin Avenue, East Lake Hi-Tech Development Zone, Wuhan, Hubei, China [After-Sales Service Unit]: Wuhan Huchuang Union Technology Co., Ltd.
A1-MT200LM Contents 1 Overview .................................................................................................................................................................3 1.1 Operation Instructions of Manual .................................................................................................................3 1.2 Overview ......................................................................................................................................................
A1-MT200LM 1 Overview 1.1 Operation Instructions of Manual 1.1.1 It is not allowed to print or disclose any content of this Manual, including pictures and audio products, under any name without the consent of Huchuang Union; 1.1.2 The equipment operator may copy some sections of this Operation Manual for internal use only, such as for instructing the user how to deal with emergencies. These sections are clearly listed in the catalogue of this manual; 1.1.3 Wuhan Huchuang Union Technology Co., Ltd.
A1-MT200LM 1.3.2 Atmospheric pressure: 70kPa~105kPa; 1.3.3 Working ambient temperature of MT200LM main unit: 0℃~+50℃; 1.3.4 Storage ambient temperature: 0℃~+50℃; 1.3.5 Relative humidity in the working environment: 0%~80% (non-condensing); 1.3.6 Power adapter (input: AC100V ~ 240V, 50/60Hz; output: 5V, 2.1A, 10.5W); 1.4 Environmental Protection Instructions 1.4.1 MT200LM contains reusable materials, and its components can be recycled after being cleaned and sterilized. 1.4.
A1-MT200LM 2 Structure Features and Equipment Parameters 2.1 Structure Features a) Charging port; a) b) USB upgrade debugging port; c) Reset port; d) SN.
A1-MT200LM e) i) f) g) h) Status indicator (red = unlocked, blue = charging, green = fully charged, purple = unlocked and charging, orange = unlocked and fully charged); j) Lock in lock-hole k) Mounting screw hole; i) j) k) 2.
A1-MT200LM Item Overall dimensions Weight Power supply mode Transmission method Parameter 83.8mm × 83.5mm × 64mm (L×W×H) 600g Powered by battery (chargeable) Wireless 1200 m in open area, 1000 m in streets, about 300 m inside Transmission distance buildings Liquid nitrogen level Detect the reaction liquid level through temperature detection Resolution 0.1℃ Accuracy ±0.
A1-MT200LM After MT200LM and MT500 devices are installed normally, users can download the "Laboratory Monitoring System" APP to view the temperature data tested; or view the temperature data through the computer web version. You can also view the corresponding data on a local PC; besides, you can perform fingerprint verification unlocking on the local PC to unlock it. 4 Installation Instructions 4.
A1-MT200LM Check whether the product is damaged. MT200LM, as a wireless transmitter, shall be equipped with a fingerprint all-in-one machine; MT200LM is restricted by transmission distance: Indoor wall-through distance is no more than 300m. 4.
A1-MT200LM b) Install the intelligent lock according to the drawing and tighten the screws to prevent the lock from being loose; Fixed support Intelligent lock Countersunk hexagon socket head cap screw Antenna c) Install the connecting plate of the lock catch according to the drawing and tighten the screws to prevent it from being loose; 10
A1-MT200LM Countersunk hexagon socket head cap screw Lock catch Countersunk hexagon socket head cap screw Connecting plate of lock catch Fixed end of lock catch Hinge Fixed end of hinge d) Install the intelligent lock and the connecting plate of the lock catch according to the drawing, align them and tighten the screws to prevent them from being loose; 11
A1-MT200LM Countersunk hexagon socket head cap screw Countersunk hexagon socket head cap screw e) Close the intelligent lock (open the lock with the emergency key before locking, then the red indicator light will be on after unlocking) and check the locking state.
A1-MT200LM Nt100 probe Nt100 probe jack g) After connecting the touch screen all-in-one device or fingerprint terminal, check whether there are the data on the numbered liquid nitrogen tank on the monitoring software of the liquid nitrogen tank. Failed Liquid nitrogen tank lock Wait for fingerprint verification...
A1-MT200LM Introduction: The first-level administrator of the "Fingerprint entry management software" grants the relevant personnel to have the operation authority of the "Liquid Nitrogen Tank Management Software"; On the "Liquid Nitrogen Tank Management Software", implement the locking /unlocking operation and level monitoring of the liquid nitrogen tank and the recording, management and monitoring of operation processes and personnel through MT200LM. 4.4.
A1-MT200LM entry (you do not need to enter this parameter because the software automatically identifies the user name based on the ID), and the password is the password filled in at the time of entry; the instructions shall be specified in the next step; 2) Main Interface There is an administrator account by default.
A1-MT200LM STEP2 Fill in user information on the interface; The role can be laboratory or system administrator. If not administrator, laboratory is selected; Whether to select "Available" determines whether to log in to the fingerprint management system; it is generally not selected for non-administrator; if selecting it, you must fill in the password for account login; STEP3 Click "Save" button.
A1-MT200LM STEP2 Click "New" button; 5) Fingerprint entry STEP1 Select the users with the fingerprint to be entered in the user list; 17
A1-MT200LM STEP2 Click "Fingerprint Registration" button; STEP3 Select the finger with fingerprint to be entered; STEP4 Click "Start false fingerprint registration" button; Note: Click "Fingerprint registration" button, put the corresponding finger on the fingerprint sensor to enter 18
A1-MT200LM the fingerprint. Pay attention to the operation information. The fingerprint of each finger needs to be entered three times; if the fingerprint is entered successfully for three times, meaning that the fingerprint is entered successfully. STEP5 View the operation information and click "Exit" button after registration. Note: Fingerprints need to be entered one by one.
A1-MT200LM Liquid nitrogen tank lock Failed Failed Nitrogen Tank Wait for fingerprint verification... Exit Liquid nitrogen tank managem ent Lock function area Log area Status of fingerprint scanner: OK; Status of serial port (3): OK; Operation time: 50s; Terminal status: OK Liquid nitrogen tank lock Failed Failed Wait for fingerprint verification...
A1-MT200LM 3) Status Area Instructions of Liquid Nitrogen Tank c a Lock b Unlock e Detailed diagram of the d Lock Unlock status area of liquid nitrogen Lock f Unlock tank is shown in the main interface.
A1-MT200LM 3) Management Menu of Liquid Nitrogen Tank Liquid nitrogen tank lock Liquid nitrogen tank management (double click) Failed STEP1 Wait for fingerprint verification...
A1-MT200LM Liquid Nitrogen Tank Management New Liquid Nitrogen Tank Management Liquid Nitrogen Tank ID Lock ID Modify Liquid Nitrogen Tank Name Delete Exit Use Sort Note: On this interface, you can manage the liquid nitrogen tank, and perform functions such as "New", "Modify", "Delete" and "Sort" 4) New Liquid Nitrogen Tank STEP1 Liquid Nitrogen Tank Management New Liquid Nitrogen Tank Management Liquid Nitrogen Tank ID Lock ID Liquid Nitrogen Tank Nam e Modify Delete Use Liquid Nitro
A1-MT200LM Liquid Nitrogen Tank Information STEP3 Lock ID After Liquid Nitrogen No. 1 Tank Tank Name completing information, click the "OK" Enable button to return to the OK Cancel management interface; Liquid Nitrogen Tank Management New Liquid Nitrogen Tank Management Liquid Nitrogen Tank ID Lock ID Delete Modify Liquid Nitrogen Tank Nam e STEP4 Exit Use Sort No. 1 Tank No. 2 Tank Select the liquid nitrogen No.
A1-MT200LM Liquid Nitrogen Tank Management New Liquid Nitrogen Tank Management Liquid Nitrogen Tank ID Lock ID Delete Modify Liquid Nitrogen Tank Nam e STEP2 Exit Use Sort No. 1 Tank Click "Delete" button; No. 2 Tank No. 3 Tank Note: If the historical data of the liquid nitrogen tank to be deleted exist, you cannot delete it. You can only modify it and do not use it.
A1-MT200LM Liquid nitrogen tank lock Liquid nitrogen tank management (double click) Failed Wait for fingerprint verification...
A1-MT200LM Liquid nitrogen tank lock Test Administrator Passing double fingerprint verification Add nitrogen Exit Liquid nitrogen tank management Lock Lock function area Delivery Unlock Warehousing Lock Delivery Unlock Warehousing Lock Delivery After about one second, the lock will be opened, and the lock information will be displayed on the interface.
A1-MT200LM Liquid nitrogen tank lock Test Administrator Passing double fingerprint verification Add nitrogen Exit STEP3 Liquid nitrogen tank management Lock Unlock Lock Unlock Lock Unlock Check the unlocking status Lock function area Delivery Warehousing Delivery Warehousing Delivery Warehousing of all tanks.
A1-MT200LM Liquid nitrogen tank lock Failed Wait for fingerprint verification... Exit STEP2 Liquid nitrogen tank management (double click) Lock Unlock Lock Unlock Lock Unlock Lock Unlock Lock Unlock Lock Unlock Lock Enter the account in the administrator login interface displayed to log in; Only administrator can view the log area.
A1-MT200LM Liquid nitrogen tank lock Failed Failed Wait for fingerprint verification...
A1-MT200LM antenna, otherwise it is easy to break the antenna, thus affecting the wireless transmission of data. Lithium Battery Maintenance (1) When the battery level is lower than 25%, it shall be charged in time. (2) Lithium batteries may explode at high temperature. Do not put the device in a high temperature environment (higher than 60 ℃). (3) After the lithium batteries are fully charged, unplug the charger in time to extend the battery life.
A1-MT200LM reception signal. Reset MT200LM and wait for 10 minutes to check whether it's normal. The battery is used up and needs to be recharged. Whether the antenna of MT200LM is damaged. Unlocking failed of equipment No temperature is detected after the temperature probe is inserted Too large transmission distance of MT500 results in poor reception signal. Unlocking software crash. The connection between PC and MT500 is abnormal.
A1-MT200LM After the MT200LM terminal is locked or unlocked, the corresponding lock status information on the monitoring software is not consistent 7 probe connection wire is normal or broken, otherwise replace the probe Insert the probe into another normal terminal device to check whether the temperature can be normally uploaded, otherwise, it means that the probe is faulty and needs to be replaced. Check whether the MT500 connected with the PC is in a normal state.
A1-MT200LM nitrogen; MT200LM is equipped with polymeric battery and shall be not allowed to contact the high temperature environment to prevent battery damage. 8 FCC Warning 15.19 Labeling requirements. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 15.21 Information to user.
A1-MT200LM -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help.