Installation Manual
B1-MT300MIX
16
Select "9. DISPLAY SERVICE" on the setting interface, and then press the "OK" button to enter
the real-time display interface of sampled data, as shown in the picture below:
Operation Instructions of Buttons:
"RUN" key: Used to update the concentration data of O
2
and CO
2
in real time;
"STOP" key: Used to pause update the concentration data of O
2
and CO
2
in real time;
"OK" key: The system returns to the main interface when this key is pressed;
"EXIT" key: The system returns to the main screen (or the system automatically returns to the main
screen after 5 minutes) when the button is pressed;
Note: This function can view the current sampled data in real time and be convenient for engineers
to debug the device. The sampled data on the main interface will be updated only when the intake is
sampled, and will be automatically locked until the next sampling.
4 Installation Instructions
4.1 Installation Method 1
Connect the pre-mixing gas pipe to MT300MIX through a gas tee;
Gas tee
Manual
ball valve
HEPA gas
filter
Pre-mix gas
COOK
incubator
Note: The gas pipe joint
shall be fastened with the
mini M-type hose clamp.