Installation Manual
B1-MT300MIX
11
UP key: No function
DOWN key: No function
OK key: No function
Menu key: The system returns to the setting
interface when this key is pressed
Interface and Operation Instructions Upon O
2
Concentration Calibration Failure
UP key: Press this button to add the value by
0.1% and press and hold this button to add the
value by 0.1% continuously
DOWN key: Press this button to reduce the
value by 0.1% and press and hold this button to
reduce the value by 0.1% continuously
OK key: The system does not save the
calibration value and directly returns to the
setting interface when this key is pressed
Menu key: The system does not save the
calibration value and directly returns to the
setting interface when this key is pressed
Interface and Operation Instructions Upon Successful O
2
Concentration Calibration
UP key: Press this button to add the value by
0.1% and press and hold this button to add the
value by 0.1% continuously
DOWN key: Press this button to reduce the
value by 0.1% and press and hold this button to
reduce the value by 0.1% continuously
OK key: The system saves the calibration
value and returns to the setting interface when
this key is pressed
Menu key: The system does not save the
calibration value and directly returns to the
setting interface when this key is pressed
Note:
1. O
2
concentration calibration setting range: From 0% to 21%;
2. Calibrate two points: 5% concentration and 21% concentration (air calibration);
3. 21% concentration has been calibrated before delivery and does not need to recalibrate. If
calibration is required, press and hold the Menu key for more than 5 seconds to enter the calibration
interface.
3.2.6 "2.SET CO2"
Select "2.SET CO2" on the setting interface, and then press the "OK" button to enter the CO
2
concentration calibration setting interface, as shown in the picture below: