Owner manual
P/R60 & P/R65 Page 12
5.2.2 ORP Probes R65
a) Disconnect the red wire at the instrument or power supply and connect it to the
milliammeter (-) black. Connect the milliammeter (+) red to the instrument or
power supply red wire output terminal.
b) For a probe with range of –500 to 500 mV: (white and green wire open). Rinse
the probe and place it in the 200 mV solution. The milliammeter should read
between 14.4 and 16.0 mA.
Rinse the probe and place it in the 600 mV solution. The meter should read
between 19.55 and 20.45 mA.
c) For a probe with range of 0 to 1000 mV: (white and green wire joined). Rinse
the probe and place it in the 200 mV solution. The milliammeter should read
between 6.4 and 8.0 mA.
Rinse the probe and place it in the 600 mV solution. The meter should read
between 12.8 and 14.4ma.
5.3 Customer Service
5.3.1 If a problem has not been resolved with the above procedures, a telephone
consultation with your AquaMetrix representative, or directly with AquaMetrix will
provide the answer.
AquaMetrix Inc. Tel: (800) 742-1413
1245 Maple Hill Court, Unit 7 Tel: (905) 954-0841
Newmarket, Ontario Fax: (905) 954-0415
L3Y 9E8 Canada Email: support@aquametrix.com
5.4 Parts and Accessories
Description
Part #
Submersion Mounting Kit for Series –8 Probes STC60L
Submersion Mounting Kit for Series –6 Probes STC60-6
Protective shroud for Series –8 Probes PROTECTOR-3
Protective shroud for Series –6 Probes PROTECTOR-6
Hot/Wet tap Ball Valve Assembly for Series –7 P60-HTC
Union Mounting Tee w/ Adaptor for Series –8 AM-MH538N9A
500 mL pH 7 Buffer Solution A35-14
4L pH 7 Buffer Solution A35-118
500 mL pH 4 Buffer Solution A35-13
4L pH 4 Buffer Solution A35-117
500 mL pH 10 Buffer Solution A35-24
4L pH 10 Buffer Solution A35-119
500 mL 200mV Buffer Solution A35-40
4L 200mV Buffer Solution A35-115
500 mL 600mV Buffer Soltion A35-41
4L 600mV Buffer Solution A35-116
Salt bridge kit for Series –4, –6, –7 & –S C35-17(K)
(Package of 3, incl. salt bridge and cell solution)
Salt bridge kit for Series –8 AM60-9765(K)
(Package of 3, incl. salt bridge and cell solution)
Junction box with terminal strip JB-1
50 ft. 5-wire Interconnect cable dressed both ends C42-5P-050