Manual

AQUAMETRIX INC.
1-800-742-1413 www.aquamerix.com
62
APPENDIX B — Default Settings
The following program settings are the default settings for the analyzer. New analyzers will have these
settings unless the setup has already been customized for your application.
Outputs
Output 1 Output 2
Input to be transmitted tFCl °C
Low setting 0.00 0.0
High setting 2.00 100.0
ON/OFF switch ON ON
Alarms
Alarm A Alarm B
Input for alarm tFCl tFCl
Alarm function HI LO
ON/OFF switch ON ON
Set-point 0.60 ppm 0.20 ppm
Differential 0.10 ppm 0.10 ppm
Unit 1E-6 (ppm) 1E-6 (ppm)
Delay 0 s 0 s
Global Units
Metric units: temperature in degrees Celsius, chlorine concentration in parts per million (ppm)
Alarm Contacts
Configured normally open
Security
Not enabled
pH and Temperature Compensation Method for Chlorine
Automatic TC using temperature input; automatic pH compensation using pH input
Membrane Compensation
Enabled
Timer Feature
Not enabled