Owner's manual
Table Of Contents
- INDEX
- Bench Test 3
- Connections 3
- Keypad System 4
- Menu - Flow Chart 4
- Run 6
- Totalizer (Flow Mode) 6
- Password 6
- Units / Mode 7
- Flume Selection 9
- Calibration - for Level 10
- Calibration - for Open Channel Flow 11
- 4-20mA Current Loop Offset 12
- Rejection Time 12
- Relay Parameters 13
- Special Functions 15
- Sensor Location - Tank Level 17
- Sensor Mounting/Location - Open Channel 20
- Enclosure Installation 23
- Error/Warning Messages 24
- Field Troubleshooting 25
- Installation Considerations In Noisy Environments 27
- Customer Service 29
- Instrument Return Procedure 29
- Appendix A - Options 30
- Data Logger 32
- RS232C Serial Output 38
- Appendix B - Applications Background 41
- Conversion Guide 42
- Specifications 43
- Calibration Worksheet - Level Mode 46
- Calibration Worksheet - Flow Mode 47
- Units/Mode 1
- Warranty 1
CUSTOMER SERVICE
If a problem cannot be resolved with the foregoing procedures or application assistance is required,
frequently a telephone consultation with your AquaMetrix representative or directly with AquaMetrix
will provide the answer. To contact the factory, call, fax or write to:
AquaMetrix Telephone: (905) 946-1064
4-30 Royal Crest Court Fax: (905) 946-8064
Markham, Ontario
Canada
L3R 9W8
INSTRUMENT RETURN PROCEDURE
If you are returning the instrument for service please enclose a written description of the problem and a
purchase order to cover the repair. If, in AquaMetrix judgement, the repair is to be covered under
warranty, there will be no charge against the purchase order. Be sure to pack the instruments adequately
because AquaMetrix cannot be responsible for shipping damage. For safety reasons, AquaMetrix cannot
accept instruments and sensors for repair that have not been thoroughtly cleaned to remove all process
material.
When you contact AquaMetrix please have the following information available:
1. Model number / Software Version
2. Serial number
3. Date of Purchase
4. Reason for return (description of fault or modification required)
5. Your name, company name, address and phone number
AquaMetrix
Ultrason
Page 29
USA
WARNING:This equipment generates, uses and can
radiate radio frequency energy and if not installed and used
in accordance with the instructions manual, may cause
interference to radio communications. It has been tested
and found to comply with the limits for a Class A
computing device pursuant to Subpart J of Part 15 of FCC
Rules, which are designed to provide reasonable protection
against such interference when operated in a commercial
environment.
Canada
This digital apparatus does not exceed the Class A limits for
radio noise emissions from digital apparatus set out in the
radio interference regulations of the Canadian Department of
Communicatons.
Le présent appareil numérique n'émet pas de bruits
radioélectriques dépassant les limites applicables aux
appareils numériques de la Classe A prescrites dans le
Réglement sur le brouillage radioélectriques éicter par le
ministére des Communications du Canada.