Owner's manual
Table Of Contents
- INDEX
- Bench Test 3
- Connections 3
- Keypad System 4
- Menu - Flow Chart 4
- Run 6
- Totalizer (Flow Mode) 6
- Password 6
- Units / Mode 7
- Flume Selection 9
- Calibration - for Level 10
- Calibration - for Open Channel Flow 11
- 4-20mA Current Loop Offset 12
- Rejection Time 12
- Relay Parameters 13
- Special Functions 15
- Sensor Location - Tank Level 17
- Sensor Mounting/Location - Open Channel 20
- Enclosure Installation 23
- Error/Warning Messages 24
- Field Troubleshooting 25
- Installation Considerations In Noisy Environments 27
- Customer Service 29
- Instrument Return Procedure 29
- Appendix A - Options 30
- Data Logger 32
- RS232C Serial Output 38
- Appendix B - Applications Background 41
- Conversion Guide 42
- Specifications 43
- Calibration Worksheet - Level Mode 46
- Calibration Worksheet - Flow Mode 47
- Units/Mode 1
- Warranty 1

Simul The Output simulation function controls the 4-20mA output, digital display
and control relays. Use it to simplify calibration of remote devices such as
chart recorders or remote displays, and to test Relay set-points. Simulation
digits are shown as a percentage of full scale. Use the Çand ÈArrow keys to
simulate
0% (4mA), 100% (20mA), and increments of .05% (.008mA).
Increments will automatically accelerate when the Çor È is continuously
pressed.
4mA ADJ
20mA ADJ
Use to fine tune the 4mA and 20mA calibration of the 4-20mA output.
Adjustment range is +1mA to –1mA in 0.002mA steps
Important
: The 4-20mA output will be forced to 4mA and 20mA respectively
during these adjustments.
New Password position cursor under digits and set new number between 00 and 99.
Com 24 48 96 192 Optional (with Data Logger) Speed of RS-232C transmission. Select 2400,
4800, 9600 or 19200 baud rate (must be set to match baud rate in
‘AquaMetrix Logger’ software and any modems used for serial
communication).
AquaMetrix
Ultrason
Page 16