Owner's manual

Model 2200C Page 10
The items that appear in the operations menu are:
• Password
• Test
• Calibration
• Temperature
• Status
• Cell Constant
• Range
• Output High
• Output Low
• Cycle On / Off
• Control Relay Setpoint
• Control Relay Deadband
• Alarm Relay High
• Alarm Relay Low
4.7 DIP Switches
4.7.1 The controls which are frequently used in the normal operation of the instrument are
all accessible on the control panel. Some switches, which are infrequently used,
are located on the back of the main circuit board.
4.7.2 The DIP switches are scanned only on power-up and every time the instrument is
taken out of the menu mode. Therefore, after changes to the DIP switch settings,
you must bring up the menu by pressing CALL, and then return the instrument to
the online mode by pressing RUN, in order for the instrument to scan the new DIP
values.
4.7.3 The following table describes the use of the 16 DIP switches:
Bank S1
Dip Switch
ON
OFF
1
°C
°F
2
Disabled
Enabled
3
Enabled
Disabled
4
Falling
Rising
5
Disabled
Enabled
6
Enabled
Disabled
7
Enabled
Disabled
8
Conductivity
Temperature
Bank S2
Dip Switch
ON
OFF
1
Enabled
Disabled
2
3
4
5