Users Manual

IFU-USA-2019-Rev.001 33
Manufacturers declaration-electromagnetic emissions
The System is intended for use in the electromagnetic environment (for home and professional healthcare) specied
below.
The customer or the user of the System should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
(for home and professional healthcare
environment)
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands
between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands
between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the System including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.