Users Manual
32 IFU-USA-2019-Rev.001
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
(for home and professional healthcare
environment)
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage
interruptions:
0 % UT; 250/300
cycle
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage
interruptions:
0 % UT; 250/300
cycle
Mains power quality should be that of a
typical home healthcare environment. If
the user of the System requires continued
operation during power mains interruptions,
it is recommended that the System be
powered from an uninterruptible power
supply or a battery.
Power frequency(50,
60 Hz) magnetic
eld
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
The System power frequency magnetic
elds should be at levels characteristic of a
typical location in a typical home healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.