Users Manual
IFU-USA-2019-Rev.001 31
Manufacturer’s declaration-electromagnetic emissions
The System is intended for use in the electromagnetic environment (for home and professional healthcare) specied
below.
The customer or the user of the System should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
(for home and professional healthcare
environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/
output lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should be that of a
typical home healthcare environment.
Surge
IEC 61000-4-5
+ 0.5kV, +1kV
line(s) to line(s)
+ 0.5kV, +1kV,+
2kV line(s) to earth
+ 0.5kV, +1kV
line(s) to line(s)
Not applicable
Mains power quality should be that of a
typical home healthcare environment.