Manual
35
Range Limits and Indication
Sensor
Hard Soft Certified Soft Hard
Type
Under Under Range Over Over
Range Range (F.S.) Range Range
DN Under
-20%
1
-20 to 0% 0 to 100% 100 to
120%
Over
120%
DI
Under
-20%
2
-20 to 0% 0 to 100% 100 to
120%
Over
120%
GI
Under
-20%
3
-20
3
to
0%
0 to 100% 100 to
120%
Over
120%
CI
NA
4
NA
4
0 to 100% 100 to
120%
Over
120%
AI
NA
5
NA
5
0 to 100% 100 to
120%
Over
120%
Notes:
1
DN units will measure to -20% (or -10 PSID, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
2
DI units will measure to -20% (or -150 PSI, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
3
GI units will measure to -20% (or -10 PSIG, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
4
CI units will measure to negative barometric pressure (typically -14.7
PSIG) and to +120% of full scale pressure. Calibration certification is for
-14.5 PSIG to F.S. (Full Scale Only).
5
AI units will measure to +120% of full scale pressure. Calibration
certification is for 0 PSIA to F.S. (Full Scale Only).
Hard Over & Under:
Display will read "OP" (Under -20% or Over 120%).
Soft Over & Under:
Display will still read out at (± 20%).