User Manual
16
When you decide to have your LFE cleaned, please be aware of the various capabilities of calibrating
LFE’s at Meriam. Every calibration includes: 6-8 calibration points [differential pressure 1”-8”H2O], a data
sheet with raw and reduced data, calibration curve, and instruction manual. The procedure numbers are
noted in (parentheses).
Calibrations NIST Traceable Certificate Accuracy
(% Reading)
( ) Clean per (A33555) and calibrate per (A35822) Working-Master X ±0.72% R
( ) Clean per (A33555) and calibrate per (A35822) Master-Master X ±0.54% R
( ) As Received calibration (A35822/A34777), clean, and calibrate [WM] X ±0.72% R
( ) As Received calibration (A35822/A34777), clean, and calibrate [MM] X ±0.54% R
( ) Accredited calibration ISO/IEC 17025 (contact Sales for details) X ±0.64% R
( ) Nuclear or safety application (A33544) X ±0.72% R
( ) Subcontracted per (A35352) X ±0.50% R
Options
( ) 2 additional calibration points beyond full scale- up to 12”H2O
( ) Additional calibration points beyond full scale at the following settings _____________________.
( ) Additional calibration points below 1”H2O at the following settings _______________________.
( ) Oxygen cleaning per (A50558)
( ) Calibrating the LFE with differential pressure transmitter
( ) Hydrostatic leak testing per (A33559)
( ) Pneumatic pressure test per (A70763)
The standard reference unit for flow rate is (cfm) cubic feet per minute. The additional units available are:
( ) Liters
( ) Cubic Centimeters
( ) Meters
( ) Pounds (include flowing temperature and pressure)
( ) Kilograms (include flowing temperature and pressure)
( ) Other _____________________
Time constants:
( ) Second
( ) Minute
( ) Hour
The standard reference for differential pressure is inches of water at 4°C. The additional units available are:
( ) Millimeters of water @ 4°C
( ) Centimeters of water @ 4°C
( ) Pascals
( ) Kilopascals
( ) Other _____________________
A second data sheet and calibration curve using the listed units will be included with the calibration