IMPORTANT: If the Firmware update procedure fails, repeat the Firmware update procedure by
pressing the “Try Again’ Button (it will bring you back to step 7)
Avoid powering down the HORUS if the firmware update has failed. It is mandatory that you
attempt to re-update the firmware, refresh the online web page if necessary. Or go back to the
previous browser page and select the firmware again and re-update it. If you do not proceed in
this way the Horus may no longer start up. Then a return to factory might be required.
If a fail message is posted saying “Wrong maintenance version, must be >= 23”. This means
that you must first install the latest Maintenance Firmware prior to updating to the latest
Firmware
Contact us at support@merging.com if your firmware update cannot be executed