2 Technical data
Model 7 1014222945
Volume [liter] 7 10 14 22 29 45
Usable bath length A [mm] 240 350 350 350 590 590
Usable bath width B [mm] 210 210 290 290 350 350
Usable bath depth C [mm] 140 140 140 220 140 220
Housing length D [mm] 468 578 578 578 818 818
Housing width E [mm] 356 356 436 436 516 516
Housing height (with flat cover) F [mm] 238 238 238 296 238 296
Housing height (with gable cover) G [mm] 337 337 347 405 343 401
Weight [kg] 11 14 16 17 24 26
Current consume [A] 5,2 5,2 7,8 8,7 10,4 12,2
Power [W] 1200 1200 1800 2000 2400 2800
Ambient conditions Ambient temperature 5ºC - 40°C
rh max. 80% (no condensation)
Overvoltage Category: II
Contamination degree: 2
Setting temperature range 10°C to 95°C
with activation of the boiling mode up to 100°C, see
section 4.3.1
Setting accuracy 0,1°C
Indication accuracy 0,1°C
Working temperature range 5ºC above ambient temperature to nominal temperature