Menu functions
7.5
Calibration
The appliances are temperature calibrated and adjusted at the factory. In case readjustment
should be necessary later on – for example due to influence of the chamber load – the ap-
pliance can be calibrated customer-specifically using three calibration temperatures of your
choice:
Cal1
Temperature calibration at low temperature
Cal2 Temperature calibration at medium temperature
Cal3 Temperature calibration at high temperature
To guarantee perfect control, we recommend to calibrate the appliance once a year.
For temperature adjustment, you will need a
calibrated reference measuring device.
CAL 1
+0,5 K
CAL 3
-0,8 K
CAL 2
+2,6 K
180°C
40°C
120°C
0°C
factory calibration
Abb. 26
Schematic example of temperature adjustment
Example:
Temperature deviation at 120 °C should be corrected.
1.
Activate the adjustment setting. To do so,
press the activation key on the right side
of the
CALIB
display. The display is en-
larged and the first calibration tempera-
ture – in this case 40 °C – automatically
highlighted.
UFP 100
Cal1
°C -0,2
K
Cal2 100.0 °C +0,1 K
Cal3 180.0 °C -0,2 K
JUSTIEREN
Temperatur
PROGRAM
TIME
USER ID
PROTOCOL
PR
OG
RAM
TIME
US
ER I
D
P
R
O
T
OCO
L
letzte Justierung 12.10.2012 12:00
40.0
C K0,2-
C K0,1+
C K0,2-
Calibration
Temperature
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
100.0
180.0
40.0
2.
Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
C K0,2-
C K0,1+
C K0,2-
JUSTIEREN
Temperature
Last adjustment 12.10.2012 12:00
Cal1
Cal2
Cal3
180.0
40.0
100.0