Owner manual

D24022 | Date 10/2014 35
Menu functions
All other settings can be made accordingly. The settings possible are described in the follow-
ing sections.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati-
cally returns to the main menu and restores the former values.
7.3
Setup
In the
SETUP
display, you can set the following parameters:
the
IP address
and
Subnet mask
of the appliance's Ethernet interface (for connection to
Subnet mask of the appliance's Ethernet interface (for connection to Subnet mask
a
network)
the
Unit
of the temperature display (°C or °F, see page 36)
the type of temperature monitoring (TWW or TWB,
Alarm Temp,
only if equipped with
a second temperature sensor, option A6) and the trigger temperature of the monitoring
function (
Max alarm
, see page 37)
how the digital backwards counter with target time setting works (
Timer mode
, see page
38)
the type of the slide-in unit (Grid or Shelf, see page 38)
Remote control (see page 38)
Gateway (see page 39)
If the SETUP menu contains more entries than can be
displayed, this is indicated by the display “1/2”. This
means that there is a second „page“ of entries.
To display the hidden entries, use the turn control
to scroll beyond the lowest entry. The page display
changes to “2/2”.
7.3.1
IP address and subnet mask
If you want to operate one ore more appliances in a
network, each appliance must have its
own unique IP address for identification. By default, each appliance is delivered with the IP
address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
192.168.1.216
Editor
Programmname
Simulation Protokoll
-
+
-
+
Test 01
180
.0°C
i
HPP 250 Labor
i
37
.0°C
44
.4%rh
STAND BY
Programm negnulletsniEllokotorPtäreG efliHnekcurD
AtmoCONTROL
Fig. 21
Operation of several appliances in a
network (schematic example)
Setup
Unit
IP adress 255.145.136.225
Subnet mask 255.255.0.0
Alarm Temp
°C F
Timer mode
TWW TWB
1/2