Owner manual

D24022 | Date 10/2014 27
Operation and control
3.
Press the confirmation key.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
The display now shows the remaining time in
a large font and the approximate end time in
a smaller font beneath.
TIMER
30m04h
End
13:30 23.11.
4.
Now, as described under
5.4.2
, set the individual values for temperature, air flap position
etc. which you want the appliance to operate at.
For universal ovens UN/UF and incubators
IN/IF, the set values can be changed while the timer elapses. The changes are effective im-
mediately. For
sterilisers SN/SF, parameters cannot be changed while the timer elapses.
For universal ovens UN/UF and incubators IN/IF, you can choose if the timer should run
setpoint-dependent or not in the
Setup
– this determines if the timer should not start
until a tolerance band around the set temperature is reached or if it should start right after
activation (see page 38). If the timer runs setpoint-dependent, this is indicated by the
symbol in the timer display.
symbol in the timer display.
When the timer has elapsed, the display shows 00h:00m. All func-
tions (heating etc.) are switched off. If a fan had been active, it will
keep on running for a short safety period. In addition, an acoustic
alarm sounds, which can be turned off by pressing the confirma-
tion key.
To deactivate the timer, open the timer display by pressing the
activation key again and then turning the turn control
to reduce
the timer setting until --:-- is displayed. Confirm with the confirma-
tion key.
5.5
Temperature monitoring
The appliance is equipped with a double overtemperature protection (mechanical/electronic)
in accordance with DIN 12 880. This serves to avoid damage to the chamber load and/or ap-
pliance in case of a malfunction:
Electronic temperature monitoring (
TWW/
TWB) (TWB only if equipped with a second tem-
TWW/TWB) (TWB only if equipped with a second tem-TWW/
perature sensor, option A6)
mechanical temperature limiter (TB)
TIMER
00m00h
End
13:30 23.11.
TIMER
--m--h
End
9:00 23.11.