Owner manual

5
Supplement to Operating Instructions UFP 600 TS/800 TS
4. Temperature control programme suggestion on
MemoryCard
Ramp.
no.
Time
tem-
pera-
ture
Fan Flap Ramp close
statement
Description
1 0:01 160 °C 100 % CLOSE SPWT(T) Quickly heating up to set
temperature. Wait until the
setpoint temperature has
been reached.
2 2:00 160 °C 100 % CLOSE NEXT Hold time
3 0:01 45 °C 100 % OPEN SPWT(T) Cooling down to unloading
temperature. If necessary,
open the air flap for faster
cooling down.
4 0:01 20 °C OFF CLOSE END Turning off the fan and close
programme
The confirmation button cannot be pressed until all doors have been closed. This way, it
is not possible to open both doors at the same time.
5. Attachment
Memory medium with PLC programme: 2011-07-08 ra UFP600TS
Function block diagram (FBD) of the PLC
Parameters/block diagram of the PLC