Operating instructions

UP/BP/ULP/SLP 400 – 800 7
model volume current power voltage ±10% weight
UP/ULP/SLP 400 53 l 6,1A 1400 W 230 V~ 35 kg
UP/ULP/SLP 500 108 l 8,7 A 2000 W 230 V~ 50 kg
UP/ULP/SLP 600 256 l 10,4 A 2400 W 230 V~ 87 kg
UP/ULP/SLP 700 416 l 5,8 A 4000 W 400 V 3ph/N~ 121 kg
UP/ULP/SLP 800 749 l 7,0 A 4800 W 400 V 3ph/N~ 164 kg
BP 400 53 l 3,5 A 800 W 230 V~ 35 kg
BP 500 108 l 3,9 A 900 W 230 V~ 50 kg
BP 600 256 l 7,0 A 1600 W 230 V~ 87 kg
BP 700 416 l 7,8 A 1800 W 230 V~ 121 kg
BP 800 749 l 8,7 A 2000 W 230 V~ 164 kg
This equipment is intended for operation on a supply system with a system impedance Z
max
at the transfer point (building connection) which does not exceed 0.12 Ohm. The user has
to ensure that the equipment is being operated on a supply system which meets this
requirement. If necessary, details of the system impedance can be obtained from the local
power supply authority.
Quality of material
Memmert is using stainless steel (Spec. 1.4301) for the external casing as well as
for the interior, an outstanding material because of its high stability, optimum
hygienic features and corrosion resistance against many (not all!) chemical
compounds (Warning against chlorine compounds, for example !)
WARNING!
Before removing top cover – pull out plug !