Operating instructions

38 UP/BP/ULP/SLP 400 - 800
14.4 Example of programming a sterilisation profile
The following example shows a possible sterilisation profile. This example uses 60 min
at a temperature of 180°C.
To programme this temperature profile, select the submenu EDIT, with the main switch
in position .
ϑ
180°C
20°C
1min
t
FAN
FLAP
set temperature
actual temperature
wait
1 h
1min
wait
1min
End
The temperature of
180°C has to be
reached in 1 minute.
The fan motor runs at
maximum speed.
The segment closing
command
SP.WT.1
ensures that the
sterilisation time begins
only when the 180°C
temperature has been
reached.
ϑ
180°C
20°C
1min
t
FAN
FLAP
set temperature
actual temperature
wait
1 h
1min
wait
1min
End
The sterilisation
time has to be
exactly 1 hour.
The fan motor runs
at maximum speed.