In particular with heavy loads, these approximate values should not be used without
proper testing. In order to ensure safe sterilisation the individual sterilisation process must
be validated, using e.g. additional temperature sensors or germ packets.
During sterilising the air flap has to be closed after the moist load has been dried.
14.3 Notes on working with sterilising cassettes:
The cassettes should be placed in the unit so that the hot air can flow freely through the air
slots (in an oven with fan the air stream flows crosswise through the working space). The
instructions about correct loading, which can be found in this operating manual and on the
unit ,must also be observed when using sterilising cassettes. This makes sure that proper
temperature distribution is achieved and that the development of cold air pockets is
prevented.
The sterilisation load must be packed in aluminium foil as shown in the table and must be
placed in the sterilising cassettes. The air slots of the cassette must be opened for
sterilising.
A temperature sensor for verifying the temperature can be placed into the sterilising
cassette through opening 2.
After the sterilising time has elapsed, the air slots must be closed by shifting knob 1 in the
direction of the arrow.
This ensures that the sterilised goods can be preserved in the closed cassette for a few
days.
1
2
WARNING!
Models 700 and 800 are fitted with a door closure mechanism
incorporating a lock. If the user, against our express warning,
enters the interior of the steriliser it is essential that he first pulls
off the key and keeps it on his person.