Operating instructions

UP/BP/ULP/SLP 400 – 800 25
The dynamic operating menu only shows as many programme segments as programmed,
i.e. until a programme segment is closed with a programme end command (LOOP, HOLD,
END). If the programme end command of the existing last programme segment is replaced
by NEXT or SP.WT. an additional programme segment is added (up to 9 segments).
Note: The return from any menu to the actual temperature is achieved by selecting the
menu item EXIT or occurs automatically after 30 sec.
In the menu item TIME the duration of the
programme segment (example: Ramp I) is
programmed.
With the parameter TEMP the final
temperature of the segment is set
(example: Ramp I).
For ovens with a fan, FAN serves to pre-
set the speed of the fan in occurs in 10 %
stages. OFF means: fan is switched off
completely.
FLAP serves to pre-set the air flap in 10%
stages.
Segment end
commands.
After selection
of the segment
end command
NEXT or
SP.WT a further
programme
segment is
shown and the
parameters can
be adjusted as
described
above.
EXIT serves to leave the EDIT submenu. By selecting NO it is possible to remain in the
submenu. If YES is selected you leave the EDIT level and the present temperature is shown.