Operating instructions

24 UP/BP/ULP/SLP 400 - 800
9.3 Editing the programme / EDIT submenu
The programming submenu is opened by selecting YES in the menu item EDIT (or VIEW
during programme run). The parameters for each programme segment (max. 9) are set
here: Duration (
time
), final temperature of the segment (
Temp
), speed of the fan
(
Fan
-if applicable) and the segment end command . The status LED edit indicates that
the controller is in the EDIT submenu.
Menu items
= turn rotary knob clockwise
= turn rotary knob anticlockwise
Use of turning the rotary knob:
Without set
button:
Together with set
button:
selection of the
menu item
TIME
DATE
YEAR
FEED
ADRES
TYPE
EXIT
press:
set point is
displayed
(temperature or
time, decimal
point is
flashing)
hold down:
displayed value
can be changed
with the rotary
knob.
release:
value is stored
in memory