accessories (Generation 2003) Owner manual

5
Supplement to Operating Instructions CTC/TTC
1.6.2 Heating up process
To accelerate the heating up process, it is recommended to set a higher chamber temperature
until the chamber load has reached the desired temperature. Afterwards, the chamber tem-
perature is also set to the desired test temperature.
Programming example:
Quickly heating up the core temperature of a test object to +20 °C. The chamber temperature
is set to +30 °C until the temperature of the test object has reached +20 °C. ASF tolerance
setting: ± 10 K.
Rampe Time Temperature Fan Rh Cond
1 0:01 +30 °C 100 % off Spwt(T)
2 0:01 +20 °C 100 % off Next
3 1:00 +20 °C 100 % off Next
+20 °C
+30 °C
+40 °C
Temperature in the test chamber
Temperature of the test object
ASF tolerance band