Owner's manual

62
Maintenance and Servicing
Fig. 36
Dismantling the right and left inner walls.
Remove the silicone plugs and 8 screws on
each side and pull out the wall.
A
Fig. 37
Pull the temperature sensors (A) out of the
holders and unscrew the test chamber rear
wall
8.3 Regular maintenance
Every three months if in permanent operation
Grease the moving parts of the doors (hinges
and lock) with thin silicone grease and check
that the hinge screws are not loose.
Check that door seals fit tightly, adjust door
if necessary (see Chapter 8.4 ).
Check that the condensation water runoff
tube ( Fig. 38 ) at the rear of the chamber
interior is not blocked and clean if necessary.
Every six months
Check air filter of cooling unit for dirt and
clean if necessary (see Chapter 8.5 ).
Annually:
Clean air filter of cooling unit (see Chapter 8.5 ).
Grease the moving parts of the doors (hinges and lock) with thin silicone grease and check
that the hinge screws are not loose.
Check that door seals fit tightly, adjust door if necessary (see Chapter 8.4 ).
Check the safety labels and renew warning stickers if necessary (see Fig. 1 on page 7 ).
Fig. 38 Condensation water runoff
tube