Manual
PERFECT
Muster
page 25
Heater power reduction
for gentle heating of the load and reduction of the mean current taken
from the power supply.
Note: reduction in heater power may lead to high temperatures no
longer being achieved.
Range: 50 to 100%
Tolerance margin ASF
Range:
Universal ovens Uxx 2 to 20
Sterilisers Sxx 2 to 20
Incubator Ixx 0.5 to 5
(see Section „Temperature monitor”)
Temperature monitor function
Adjustable temperature monitor (TWW)
Protection Class 3.1 to DIN 12 880
Adjustable temperature limiter (TWB)
Protection Class 2 to DIN 12 880
(see Section „Temperature monitor”)
Language
Settings: GERMAN, English, franCAIS, ESPANOL and italIANO
Calibration correction for user-calibration
CAL1 to CAL3
ADJUST – TEMPERATURE CALIBRATION
READJUST – TEMPERATURE CORRECTION
(see Section „Calibration“)
Exit setup mode
= store all settings and exit SETUP mode.
12.1 Real-time clock
The real-time clock is set in SETUP and includes date and clock time.
The real-time clock serves for documentation according to GLP.
Date and clock time are marked in the report print.
On the graphics print the time axis is marked in real-time.
The clock runs with a buffer battery independently of the mains power supply.
The built-in lithium battery Type CR 2032 has a life of approx. 10 years.