Example: Humidity deviation at 80 % should be corrected.
1.
Set humidity balance point in the SETUP to
RH 80
and set the accompanying compensation
correction value to
0.0 %
rh:
PRINT
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO
mb
2
IN 1
IN 2
OUT
MIN
AUTO
MAX
rh
mb
%
2.
With a calibrated reference instrument, an actual humidity of 78 % rh is measured at
normal operation, with a defined setpoint humidity of
80 %
rh
.
3.
Set the compensation correction value in the SETUP for
RH 80
to
–2.0 %
rh
:
PRINT
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa Su
3
4
2
1
STERI DEFRO
°C
°C
rh
mb
%
CO
mb
2
IN 2
OUT
IN 1
IN 2
OUT
MIN
AUTO
MAX
rh
mb
%
IN 2
OUT
4.
After the calibration procedure, the reference instrument should display 80.0 %.
With
RH 20
a further comparison can be programmed at 20 % relative humidity.
If all
compensation correction values are set to
0.0
% rh, the factory calibration settings
are restored.
7.8.3
CO
2
calibration
Customer calibration of appliance on controller by means of three CO
2
points:
CO2 5
CO
2
calibration at 5 % CO
2
content
CO2 10
CO
2
calibration at 10 % CO
2
content
CO2 15
CO
2
calibration at 15 % CO
2
content
For each selected balance point, a positive or negative compensation correction value can be
set.
co2 5
-0,4 %
20%
0%
10%
co2 10
+1,5 %
co2 15
-0,7 %
Default calibration
Fig. 30
CO
2
calibration (example)