Manual
57
Advanced Functions
7.4
Communication interfaces
7.4.1
USB interface
The cooled incubator is equipped by default with a
USB interface. In accordance with USB
specification (see
Fig. 5
on page
14
). With this interface, it is possible to control and log the
appliance remotely from the computer. This is done with the help of the
“Celsius“ software.
To do this, the cooled incubator must be given a unique
device address in the SETUP sub-
menu, menu item
ADDRESS
, via which the computer communicates with the appliance (see
page
52
). Using this, the respective cooled incubator can be selected and programmed from
the computer. The default setting is
ADDRess 0
.
If several appliances are to be connected to a computer via
USB interface, an appropriate
interface on the computer and a separate cable are required for each. The maximum
cable
length is 5 m.
7.4.2
Ethernet interface
The cooled incubator can optionally be equipped with an
ethernet interface to connect it to
a network. For identification purposes, each cooled incubator connected must have its own
unique IP address. Each appliance is delivered by default with the IP address 192.168.100.100.
The programme "
XTADMIN", which can be found on the “Celsius“ CD-ROM, can be used to
change the IP address.
How to set the IP address is described in the enclosed ”Celsius“ manual.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
Fig. 26
One or more cooled incubators are connected to a network via
ethernet interface
(schematic diagram)
Fig. 26
(schematic diagram)
Fig. 26