Manual

53
Advanced Functions
Tolerance band ASF
Adjustment range:
0.5
to
5°C
(see page
53
)
Language
Language
Setting:
GERMAN, English, franCAIS, ESPANOL
and
italIANO
Corrective values
CAL 1-3, rh20, RH80, CO2 5, CO2 10, CO2 15
for customer-side
calibration of temperature, humidity and
CO
2
(see chapter "
Calibration
" on page
59
)
Leave setup
Leave setup
Save all settings and leave the SETUP mode
The
realtime clock which is set in the SETUP contains the date and clock time. It is used
for logging purposes in accordance with GLP. Date and clock time are specified on the log
printout. On graphical printouts, the time axis is labelled with the realtime. The clock is
battery-buffered and is independent of the mains connection. The integrated lithium battery
of the type CR 2032 has a lifetime of approx. 10 years.
7.3
Temperature monitoring
The
monitoring temperature is measured via a separate
Pt100
temperature sensor in the
chamber interior. The monitoring unit is used to protect the chamber load and as a protection
for the appliance and surroundings.
The appliance is equipped with a double overtemperature protection (mechanical/electronic)
in accordance with DIN 12880.
°C
MIN
AUTO
MAX
1
2
3
1
6
5
4
Fig. 22
Temperature monitoring display
1
Visual
Fig. 22
Visual
Fig. 22
alarm icon lit up: TB
Temperature monitoring display
alarm icon lit up: TB
Temperature monitoring display
alarm flashing: TWW
Temperature monitoring display
alarm flashing: TWW
Temperature monitoring display
alarm, ASF
alarm
2
Undertemperature protection
alarm icon lit up: TB
Undertemperature protection
alarm icon lit up: TB
alarm flashing: TWW
Undertemperature protection
alarm flashing: TWW
3
Automatic temperature monitor (ASF, see page
56
)
56 )56
4
Overtemperature protection (TWW, TWB, see page
54
)
54 )54
5
Trigger temperature
6
Acoustic
alarm icon
Trigger temperature
alarm icon
Trigger temperature