Operation and Control
21.
Setting the relative humidity of the third ramp segment:
Turn the push-turn control to the right until the humidity
display flashes
.
Hold down the SET key and set the desired humidity
setpoint to
50.0 % rh
with the push-turn control.
%
rh
22.
Setting the CO
2
content of the second ramp segment
(only for appliances with CO
2
supply, otherwise please
continue with point 23):
Turn the push-turn control to the right until the CO
2
display flashes.
Hold down the SET key and set the desired CO
2
setpoint
to
0.0%
with the push-turn control.
CO
2
23.
Setting the lighting of the second ramp segment (only
for models with lighting, otherwise please continue with
point 24):
Turn the push-turn control to the left until the
lamp dl
display appears.
Hold down the SET key and set to
on
with the push-turn
control.
24.
Setting the close statement of the third ramp segment:
Turn push-turn control to the right until a close state-
ment appears in the display, e.g.
end,
and press the SET
key briefly to confirm.
25.
Leave programming write mode EDIT
Turn push-turn control to the right until
EXIT
appears in
EXIT appears in EXIT
the display, and press the SET key briefly to confirm.
26.
Setting the temperature monitoring:
Turn the push-turn control to the right and adjust the
temperature monitoring (for more detailed information
on this, see page
53
).
°C
MIN
AUTO
MAX
27.
Activating the programme:
Turn the push-turn control to the right until the stop
icon
flashes.
flashes.
Hold down the SET key and select the start icon
with
with
the push-turn control. Release the SET key, and the pro-
gramme is activated.
5.10.8
Operation with
PC/
laptop (optional)
The cooled incubator can optionally be used, controlled and programmed with a PC/laptop.
For this purpose, it is equipped with a USB or
ethernet interface on the rear side (see page
14
and
57
).
The control of the appliance with the Memmert computer software “Celsius“ is
described in a separate manual.