Operation and Control
14.
Setting the fan speed of the second ramp segment:
Turn the push-turn control to the right until the fan
display flashes.
Hold down the SET key and set the desired fan speed of
50 % (five bars lit up) with the push-turn control.
15.
Setting the relative humidity of the second ramp seg-
ment:
Turn the push-turn control to the right until the humid-
ity display flashes
.
Hold down the SET key and set the desired humidity
setpoint to
70.0 % rh
with the push-turn control.
rh
%
16.
Setting the CO
2
content of the second ramp segment
(only for appliances with CO
2
supply, otherwise please
continue with point 17):
Turn the push-turn control to the right until the CO
2
display flashes.
Hold down the SET key and set the desired CO
2
setpoint
to
15.0%
with the push-turn control.
CO
2
17.
Set lighting of the second ramp segment (only for mod-
els with lighting, otherwise please continue with point
18):
Turn the push-turn control to the left until the
lamp uv
display appears.
Hold down the SET key and choose
on
with the push-
turn control.
Turn the push-turn control to the left until the
lamp dl
display appears.
Hold down the SET key and set to
On
with the push-turn
control.
18.
Setting the close statement of the second ramp seg-
ment:
Turn the push-turn control to the right until a segment
close statement, e.g.
end, appears.
Hold down the SET key and set the close statement
next
with the push/turn control.
19.
Setting the duration of the third ramp segment:
Select the time display with the push-turn control.
Hold down the SET key and set the time to 1.00 with the
push-turn control.
h
20.
Setting the temperature of the third ramp segment:
Turn the push-turn control to the right until the tem-
perature display flashes.
Hold down the SET key and set to
20.0 °C
with the push-
turn control.
°C