Operation and Control
7.
Setting the fan speed of the first ramp segment:
Turn the push-turn control to the right until the fan
display flashes.
Hold down the SET key and set the desired fan speed of
30 % (three bars lit up) with the push-turn control.
set
off
on
push
card
4
3
2
1
mb
CO
mb
2
IN 1
IN 2
OUT
IN 1
IN 2
OUT
PRINT
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa Su
°C
°C
MIN
AUTO
MAX
h
rh
DEFROSTERI
%
1 2 3 4 5 6 7 8 9
10111314151617
8.
Setting the relative humidity of the first ramp segment:
Turn the push-turn control to the right until the humidity
display flashes
.
Hold down the SET key and set the desired humidity
setpoint to
70.0 % rh
with the push-turn control.
%
rh
9.
Setting the CO
2
content of the first ramp segment (only
for appliances with CO
2
supply, otherwise please con-
tinue with point 10):
Turn the push-turn control to the right until the CO
2
display flashes.
Hold down the SET key and set the desired CO
2
setpoint
to
5.0%
with the push-turn control.
CO
2
10.
Set lighting of the first ramp segment (only for models
with lighting, otherwise please continue with item11):
Turn the push-turn control to the left until the
lamp dl
display appears.
Hold down the SET key and choose
on
with the push-
turn control.
11.
Setting the close statement of the first ramp segment:
Turn the push/turn control to the right until a segment
close statement, e.g.
end,
appears.
Hold down the SET key and set the close statement
SPWT [TH]
with the push/turn control.
12.
Setting the duration of the second ramp segment:
Turn the push-turn control further to the right until the
time display flashes.
Hold down the SET key and set time to
0.45
with the
push-turn control.
h
13.
Setting the temperature of the second ramp segment:
Turn the push-turn control to the right until the tem-
perature display flashes.
Hold down the SET key and set the desired temperature
setpoint to
37.0
°C
with the push-turn control.
°C