Manual

19
Design and Function
D
A
E
B
F
C
Fig. 7
Dimensions of ICH cooled incubators
2.11
Ambient conditions
The cooled incubator may only be used in enclosed rooms and under the following ambient
conditions:
Ambient temperature: 5 ºC to 28 ºC
Humidity: max. 80 %
Degree of pollution: 2
Altitude of installation: max. 2000 m above sea level
The cooled incubator may not be used in areas where there is a risk of explosions. The
ambient air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The
cooled incubator is not explosion-proof.
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the interior and, as a consequence, could result in short circuits or dam-
age to electrical parts. For this reason, sufficient measures are to be taken to prevent large
clouds of dust or aggressive vapours from developing.
2.12
Accessories
One sliding shelf
Water supply tank and connection tube
Additional for cooled incubators with CO
2
2
supply:
supply:
Gas pressure tube