Manual

14
Design and Function
2.5
Electrical equipment
Operating voltage: See nameplate (page 16), 50/60 Hz
Current consumption: See nameplate (page
17
)
Protection class 1, i.e. operating insulation with PE conductor in accordance with EN
61010
Protection type IP 20 acc. to EN 60 529
Interference-suppressed acc. to EN 55011 class B
Appliance fuse: Fusible link 250 V/15 A quick-blow
The temperature controller is protected with a miniature fuse 100 mA (200 mA at 115 V)
The CO
2
controller is equipped with a 6.3 A miniature fuse.
2.6
Connections
1
3
4
5
6
2
CO
2
Fig. 5
Connections on rear of appliance
1
Mains lead
Fig. 5
Mains lead
Fig. 5
Connections on rear of appliance
Mains lead
Connections on rear of appliance
2
CO
2
connection (only for models with CO
2
supply)
3
USB connection (optional
2
USB connection (optional
2
connection (only for models with CO
USB connection (optional
connection (only for models with CO
ethernet connection, see page
connection (only for models with CO
ethernet connection, see page
connection (only for models with CO
2
ethernet connection, see page
2
supply)
ethernet connection, see page
supply)
57
)
57 )57
4
Printer connection (see page
51
)
5
Water connection (see page
23
Water connection (see page 23Water connection (see page
)
23 )23
6
Fresh air connection (only for models with CO
Water connection (see page
Fresh air connection (only for models with CO
Water connection (see page
2
supply)
2.6.1
Electrical connection
Observe the country-specific regulations when connecting (e. g. in Germany DIN VDE 0100
with residual current circuit breaker).
This appliance is intended for operation on an electrical power system with a system imped-
ance Z
max
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
must ensure that the cooled incubator is operated only on an electrical power system that
meets these requirements. If necessary, you can ask your local energy supply company what
the system impedance is.