Manual

13
Design and Function
2.3
Configuration
2.3.1
Basic configuration
Electronic fuzzy-supported PID process controller with pulse package control and per-
manent performance adjustments and time-saving self-diagnosis system for quick error
location
All-round large-area heating with additional thermal conduction layer
Capacitive humidity sensor
Homogeneous atmosphere and temperature distribution through encapsulated, turbu-
lence-free ventilation system
Language selection
Alphanumeric text display
Integrated week time switch with group function (e.g. all workdays)
Recessing push/turn control for simple operation of the appliance
Two separate Pt100 temperature sensors DIN class A in a 4-wire circuit for control and
monitoring
Digital monitoring control for overtemperature, undertemperature and automatic setpoint
following (ASF)
Mechanical temperature limiter (TB protection class 1)
Monitor relay to switch off heating in case of fault
Visual alarm indication
Acoustic warning signal
Calibration of temperature, humidity and (optional) CO
2
at the appliance, without the
need of a separate computer
2.3.2
Additional fittings
The ICH 256 cooled incubator can be equipped with the following fittings:
CO
2
supply
Illumination simulation daylight
Illumination simulation daylight and UV light
2.4
Material
For the outer housing, Memmert deploys stainless steel (Mat.No. 1.4016) and for the interior,
stainless steel (Mat.No. 1.4301) is used, which stands out through its high stability, optimal
hygienic properties and corrosion-resistance towards many (but not all!) chemical compounds
(caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with
the materials mentioned.
A material resistance table can be requested from the manufacturer.