User Manual

D24030 | Date 10/2014 47
Menu mode
7.3.4
Timer mode
Here, you can choose whether the digital backwards counter with
target time setting (see page 29) should run setpoint-de-
pendent or not. This determines whether the timer should not
start until a tolerance band of ±3 K around the set temperature
is reached (Fig. 25, B) or if it should start right after activation
(A).
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 25
Timer Mode
A Timer independent of setpoint: Timer starts right after activation
Fig. 25
A Timer independent of setpoint: Timer starts right after activation
Fig. 25
B Timer setpoint-dependent: Timer does not start until tolerance band is reached
A Timer independent of setpoint: Timer starts right after activation
B Timer setpoint-dependent: Timer does not start until tolerance band is reached
A Timer independent of setpoint: Timer starts right after activation
7.3.5
Type of the
slide-in unit (
Grid
or
Shelf
)
Here, you have to set the type of the slide-in unit (grid or shelf)
used. The selection Shelf enables you to adjust the control
function to the different air flow characteristics in the interior
when using optional sliding shelves instead of the grids that are
part of the standard delivery.
SETUP
Unit
IP address 255.145.136.225
Subnet mask 255.255.0.0
Alarm temp
°C F
TWW TWB
Timer mode
Slide-in unit
Grid Shelf
SETUP
Unit
IP adress 255.145.136.
225
Subnet mask 255.255.0.0
°C °F
Timer mode
Alarm temp
TWW TWB
Slide-in unit
Grid Shelf