User Manual

D24030 | Date 10/2014 41
Malfunctions, warning and error messages
Error description Cause of error Rectifying errors See
Display T:E-3 in the
temperature display
TEMP
T:E-3 Set 37.0 °C
37.4°C
Temperature operating
sensor defective. The moni-
toring sensor takes over the
measurement function.
The appliance can
be temporarily
operated
Contact customer
service as soon as
possible
page 2
Error message AI E-3 in
the temperature display
TEMP
AI E-3 Set 37.0 °C
37.4°C
Temperature monitoring
sensor defective. The ope-
rating sensor takes over the
measurement function.
The appliance can
temporarily be kept
in service
Contact customer
service as soon as
possible
page 2
Error message E-3 in
the temperature display
TEMP
Set 45.0 °C
E-3 °C
Operating and monitoring
sensor defective
Switch off appli-
ance.
Remove the cham-
ber load
Contact customer
service
page 2
Error message E-6 in
the humidity display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective
No humidity control
possible
Contact customer
service
page 2
When switching on
the appliance, the start
animation is displayed
in another colour than
white
Cyan
:
Not enough storage
space on the SD card
Red
:
The system files could
not be loaded
Orange
: The
fonts and images could
not be loaded
Contact customer
service
Contact customer
service
Download the firmware
update at memmert.
com and install it
page 2