User Manual

34
D24030 | Date 10/2014
Operation and control
Automatic
temperature monitor (
temperature monitor (
ASF)
ASF)
temperature monitor (ASF) temperature monitor (
temperature monitor (ASF) temperature monitor (
ASF is a monitoring device that automatically follows the set temperature setpoint within an
adjustable tolerance band (Fig. 20).
The ASF – if switched on – is automatically activated as soon as the actual temperature value
reaches 50 % of the set tolerance band of the setpoint (in the example: 50 °C
±
1 K) for the
first time (section A).
When the temperature violates the set tolerance band around the setpoint (in the example in
Fig. 20:
50 °C ± 2 K) – e.g. if the door is opened during operation (section B of illustration) – the
alarm
is set off. The ASF
alarm is automatically triggered as soon as 50 % of the set tolerance band
of the setpoint (in the example: 50 °C ± 1 K) are reached again (section C).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in this
example: The setpoint is changed from 50 °C to 25 °C, section D) until the tolerance range of
the new temperature setpoint is reached again (section E).
52°C
48°C
ASF active
52°C
48°C
ASF alarm
ASF active ASF active
27°C
23°C
°C
50°C
25°C
t
A B C D E
AUTO AUTO AUTO
Fig. 20
Schematic diagram of how the ASF temperature monitoring works
Adjusting temperature monitoring
Adjusting temperature monitoring
For cooled incubators
IPP
PLUS
:
Press the activation key to the left of the
ALARM
display. The min setting (under-
temperature protection) is automatically
activated.
Continue with point 1.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
180.0 °C
Set 999.9 °C
TEMP
TIMER
End 14:45
FLAP
Holz trocknen
aufheizen
09:12h
Fr 20.10.2010 20:31
min
000°C
ALARM
0 12
%rh°C
DRAPH
off
ON
44h:44m
20%
ON
max
000°C
auto off
99K
-
+
ALARM
0.0K
min
°C
auto
max
40.0 °C
15 0.
+
-
For constant climate chambers HPP:
Press the activation key to the left of the
ALARM
display. The temperature monitoring
setting is automatically activated (
).
).
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
180.0 °C
Set 999.9 °C
TEMP
TIMER
End 14:45
FLAP
Holz trocknen
aufheizen
09:12h
Fr 20.10.2010 20:31
min
000°C
ALARM
0 12
%rh°C
DRAPH
off
ON
44h:44m
20%
ON
max
000°C
auto off
99K
-
+
ALARM
max
40.0°C
auto
0.0K
min
15.0°C
+
-