HPPlife SUPPLEMENT TO THE OPERATING INSTRUCTIONS CONSTANT CLIMATE CHAMBER HPP ANIMAL KEEPING CHAMBER HPPlife 100% ATMOSAFE. MADE IN GERMANY. www.memmert.com | www.atmosafe .
Manufacturer and customer service Memmert GmbH + Co. KG Willi-Memmert-Straße 90–96 D-91186 Büchenbach Deutschland Phone: +49 (0)9122 925-0 Fax: +49 (0)9122 14585 E-mail: sales@memmert.com Internet: www.memmert.com Customer service: Service hotline: +49 (0)9171 9792 911 Service fax: +49 (0)9171 9792 979 E-mail: service@memmert.com When contacting customer service, always quote the product serial number on the nameplate. Shipping address for repairs: Memmert GmbH + Co. KG Kundenservice Willi-Memmert-Str.
About this manual About this manual Purpose and target audience This is a supplement to the operating instructions for the constant climate chamber HPP. It describes the special functions of the animal keeping chamber HPPlife and is intended for use by trained personnel of the owner, who have the task of operating and/or maintaining the respective appliance.
Contents Contents 1. Safety regulations 5 1.1 1.1.1 1.1.2 1.2 1.3 1.4 1.5 1.6 1.7 1.8 Terms and signs used........................................................................................................... 5 Terms used ........................................................................................................................ 5 Signs used .........................................................................................................................
Safety regulations 1. Safety regulations 1.1 Terms and signs used In this manual and on the appliance itself, certain common terms and signs are used to warn you of possible dangers or to give you hints that are important in avoiding injury or damage. Observe and follow these notes and regulations to avoid accidents and damage. These terms and signs are explained below. 1.1.
Safety regulations 1.2 Product safety and dangers The appliances described in this manual are technically sophisticated, manufactured using high-quality materials and subject to many hours of testing in the factory. They reflect the state of the art and comply with recognised technical safety regulations. However, there are still risks involved, even when the appliances are used as intended. These are described below. Warning! After removing covers, live parts may be exposed.
Safety regulations ► is responsible for ensuring that the maintenance plan is adhered to and that maintenance work is carried out properly (see operating instructions constant climate chamber HPP); ► has to ensure that the appliance and its surroundings are kept clean and tidy, for example through corresponding instructions and inspections; ► is responsible for ensuring that personal protective clothing is worn by operating personnel, e.g. work clothes, safety shoes and protective gloves. 1.
Description and function 2. Description and function 2.1 Description The appliance can heat the interior up to 35 ºC and cool it down to 0 ºC. In addition to the HPP standard equipment, the appliance features an exhaust air unit in the back of the chamber and 4 swivel-type LED light strips for adjusting the interior lighting in steps of 1%. In addition, the temperature-humidity working range (see chapter 2.2) can be enlarged by connecting it to a central compressed air supply (see page 11). 2.
Description and function 2.3 Technical data Appliance width [mm] Appliance height [mm] Appliance depth [mm] Depth of door lock [mm] Appliance depth (including door handle) [mm] Chamber width [mm] Chamber height [mm] Chamber depth [mm] Chamber volume [litres] Weight (in the box) [kg] Power [W] 230 V, 50/60 Hz Current consumption [A] 115 V, 50/60 Hz max. number of sliding shelves max. load per sliding shelf [kg] max.
Putting into operation 3. Putting into operation 3.1 Assemble the exhaust air unit Warning! Danger to life due to electric shock. When assembling the exhaust air unit, the chamber must not be connected to the power supply. 1. Plug the connector of the exhaust air unit into the cable socket on the back of the appliance (Fig. 3, 1). Stow away the cable in the housing of the exhaust air unit (2). 2. Insert the exhaust air unit with the pipe socket into the entry port in the back of the chamber (Fig.
Putting into operation 3.2 Connecting the appliance Connect the appliance to the energy and water supply as described in the operating manual for the constant climate chamber HPP. Connection to a central compressed air supply By doing so, the temperature-humidity working range of the appliance can be extended. This makes dehumidification of the air in the appliance possible even at 10 ºC down to 10 % relative humidity (also see page 8).
Operation and control 4. Operation and control The appliance may only be used in accordance with the applicable national regulations and laws on keeping experimental animals and only be operated by persons who have the required qualifications for working with experimental animals. This chapter merely describes the special functions of the HPPlife appliance.
Operation and control 4.2 Adjustment options You can set the following parameters after pressing the corresponding activation key (in any sequence): Temperature Adjustment range: 0 to 40 °C Heating operation is indicated by the Cooling is indicated by the symbol. TEMP symbol. 2222.4 °C Set Humidity Adjustment range: 10 to 90 % rh Humidification is indicated by the symbol. Dehumidification is indicated by the symbol.
1 2 3 4 5 6 7 Day FAN 4 Evening Night 5 Simulates the day/night cycle in the AtmoCONTROL programme software (example) FAN 3 Jump target for the daily repetition of the programme when reaching LOOP (7) 55 % relative humidity during the entire time (∞) Day simulation: 60 % fan speed, 60 % LED lighting and 22°C for 12 hours Evening simulation: 20 % fan speed, 20 % LED lighting and 22°C for 30 minutes Night simulation: 0 % LED lighting and 22°C for 11 hours with unchanged fan speed of 20 % Morning
Animal Keeping Chamber HPPlife D32802 | Date 10/2014 englisch Memmert GmbH + Co. KG Willi-Memmert-Straße 90-96 | D-91186 Büchenbach Tel. +49 9122 925-0 | Fax +49 9122 14585 E-Mail: sales@memmert.com facebook.com/memmert.family Die Experten-Plattform: www.atmosafe.