Owner's manual
HCP page 8
Model HCP 108 HCP 153 HCP 246
Chamber width A [mm] 560 480 640
Chamber height B [mm] 480 640 640
Chamber depth C [mm] 400 500 600
Housing width D [mm] 710 630 790
Housing height E (variable through adjustable feet) [mm] 778 920 938
Housing depth F (including door handle) [mm] 590 690 790
Chamber volume [litre] 108 153 246
Weight [kg] 70 90 110
Power [W] 1000 1500 2000
Max. number of shelves
577
Max. load per shelf [kg] 30 30 30
Max. load per oven [kg] 60 60 80
Ambient
conditions
• ambient temperature 5°C to 35°C
• rh 80% max. (no condensation)
• overvoltage category: II, contamination level: 2
Temperature • temperature is measued with Pt100 in 4-wire circuit
• adjustment range:: - normal operation with humidity: 20°C to 90°C
- normal operation without humidity: 20°C to 160°C
- sterilisation operation: 160°C (4 hours) via STERICard
• setting accuracy: 0,1°C
• control range: from 8°C above ambient up to 90°C / 160°C
• variation (time): ±0,1°C max
• uniformity (space): ±0.25°C max. at 50°C and 20-90%rh
±0,5°C at 90°C and 20-90%rh
±1,3°C at 160°C and with humidity control deactivated
Humidity The relative humidity in the chamber is measured by a capacitive humidity sensor and
indicated digitally in %.
• accuracy of humidity sensor: 1% rh
• adjustment range: 20 to 95%rh
• setting accuracy: 1rh%
• indication range: 10 to 98rh%
• variation (time): ±1%rh max
supply cable serial and parallel interface
H
2
O
inner
glas door
mm (minimum height)
adjusting screw
3 Technical data